Frequently Asked Questions (FAQ)
Buckys
B1: How do I connect the bucky wires to a GX/MP or AP/ATC/MPX/APX generator?
B2: How do I connect the bucky wires to a CM series generator?
B3: Why do the grids of both buckys begin to oscillate without command, after the CM series generator is powered on?
| B1: How do I connect the bucky wires to a GX/MP or AP/ATC/MPX/APX generator? | |
| Please refer to the Bucky Connections and Service Guide document. |
| B2: How do I connect the bucky wires to a CM series generator? | |
| Please refer to the CPI CMP, Millenia and Indico Bucky Interface Guide in the CPI Technical Documents Section of our website to be informed of the connections. |
| B3: Why do the grids of both buckys begin to oscillate without command, after the CM series generator is powered on? | |
| The H.V. auxiliary board has a snubber network for each bucky drive to preserve the relay contacts from high potential arcs. The snubber networks are not needed for the buckys we use. To disable the snubber networks, clip resistors R2 and R15 on the board. |